Data Foundry

Medical device recalls 2020

Total Bilirubin Reagent — Class II medical device recall Z-1426-2020

Terminated recall by Beckman Coulter Inc, reported 2020-03-11, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol)

Recall numberZ-1426-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc, Carlsbad, CA, United States
Recall initiated2019-08-15
Classified2020-03-02
Reported by FDA2020-03-11
Terminated2024-02-28
DistributedAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAE, AU, BD, BE, BG, CA, CH, CN, CW, CZ, DE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, IE, IN, IT, JP, KR, KW, LB, LT, MX, MY, NL, OM, PA, PH, PL, PT, PY, QA, RO, RS, RU, SA, SG, TH, TN, TR, TW, VN, ZA
Quantity25,383 reagents
UPC / GTIN / UDI-DI15099590233075, 15099590625597
Model numbers442745

Product

Total Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300 tests/cartridge) and 476861 (400 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233075 (300 tests/cartridge) and 15099590625597 (400 tests/cartridge)

Reason for recall

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.