Seca 402/403 Baby Scale Cart — Class II medical device recall Z-1429-2020
Ongoing recall by Seca, reported 2020-03-11, distributed in AK, AL, AZ, CA, CO, DC, DE, FL…. It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, d
| Recall number | Z-1429-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Seca, Hamburg, N/A, Germany |
| Recall initiated | 2019-06-19 |
| Classified | 2020-03-03 |
| Reported by FDA | 2020-03-11 |
| Distributed | AK, AL, AZ, CA, CO, DC, DE, FL, HI, IA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV |
| Countries outside the US | CA |
| Quantity | 254 scale |
| Reason classes | device malfunction |
| Serial numbers | 6002142000002, 6002142001101 |
| Model numbers | 4020000009, 4030000009 |
Product
Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009
Reason for recall
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1429-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.