Data Foundry

Medical device recalls 2020

Seca 402/403 Baby Scale Cart — Class II medical device recall Z-1429-2020

Ongoing recall by Seca, reported 2020-03-11, distributed in AK, AL, AZ, CA, CO, DC, DE, FL…. It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, d

Recall numberZ-1429-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSeca, Hamburg, N/A, Germany
Recall initiated2019-06-19
Classified2020-03-03
Reported by FDA2020-03-11
DistributedAK, AL, AZ, CA, CO, DC, DE, FL, HI, IA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV
Countries outside the USCA
Quantity254 scale
Reason classesdevice malfunction
Serial numbers6002142000002, 6002142001101
Model numbers4020000009, 4030000009

Product

Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009

Reason for recall

It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1429-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.