Data Foundry

Medical device recalls 2020

bellavista 1000 ventilator — Class II medical device recall Z-1439-2020

Terminated recall by Vyaire Medical, reported 2020-03-11, distributed nationwide. The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority ala

Recall numberZ-1439-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVyaire Medical, Mettawa, IL, United States
Recall initiated2019-11-25
Classified2020-03-04
Reported by FDA2020-03-11
Terminated2023-08-08
DistributedNationwide (US)
Quantity116
Serial numbersMB170104, MB170118, MB170151, MB170202, MB230000, MB230020, MB230024, MB230027, MB230031, MB230033, MB230035, MB230036, MB230037, MB230038, MB230039, MB230040, MB230041, MB230046, MB230048, MB230049, MB230052, MB230053, MB230054, MB230055, MB230056 and 91 more
Model numbers301.100.030

Product

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Reason for recall

The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1439-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.