Master Drug Library Software version 8 — Class II medical device recall Z-1451-2015

Terminated recall by Baxter Healthcare Corp, reported 2015-04-22, distributed in CA, CT, FL, IA, IL, IN, LA, MA…. Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during program

Recall numberZ-1451-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corp, Deerfield, IL, United States
Recall initiated2015-03-02
Classified2015-04-16
Reported by FDA2015-04-22
Terminated2018-05-22
DistributedCA, CT, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NH, NY, PR, SD, TX, UT, WV
Countries outside the USCA
Quantity58 mdls

Product

Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Product Code 35700BAX2 The SIGMA Spectrum Infusion Pump with Master Drug Library (MDL) is intended to be used for the controlled administration of fluids.

Reason for recall

Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.