Data Foundry

Medical device recalls 2015

Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use — Class II medical device recall Z-1452-2015

Terminated recall by Integra LifeSciences Corp., reported 2015-04-22. A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified a

Recall numberZ-1452-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corp., Plainsboro, NJ, United States
Recall initiated2015-03-13
Classified2015-04-16
Reported by FDA2015-04-22
Terminated2016-04-13
Countries outside the USCH, ES, FR, GB, GR, IT, TR
Reason classesspecification failure
Lot numbers10500028786, 105000288118, 105N00288965
Model numbersFWD301

Product

Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.

Reason for recall

A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and force vs. displacement (force needed for product to be extruded from the syringe).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1452-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.