Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use — Class II medical device recall Z-1452-2015
Terminated recall by Integra LifeSciences Corp., reported 2015-04-22. A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified a
| Recall number | Z-1452-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Integra LifeSciences Corp., Plainsboro, NJ, United States |
| Recall initiated | 2015-03-13 |
| Classified | 2015-04-16 |
| Reported by FDA | 2015-04-22 |
| Terminated | 2016-04-13 |
| Countries outside the US | CH, ES, FR, GB, GR, IT, TR |
| Reason classes | specification failure |
| Lot numbers | 10500028786, 105000288118, 105N00288965 |
| Model numbers | FWD301 |
Product
Integra¿ Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
Reason for recall
A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and force vs. displacement (force needed for product to be extruded from the syringe).