Draeger Infinity CentralStation — Class II medical device recall Z-1471-2023

Ongoing recall by Draeger Medical Systems, Inc., reported 2023-05-03, distributed nationwide. The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail

Recall numberZ-1471-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDraeger Medical Systems, Inc., Andover, MA, United States
Recall initiated2023-04-05
Classified2023-04-27
Reported by FDA2023-05-03
DistributedNationwide (US)
Quantity6,803 systems
Reason classessoftware
UPC / GTIN / UDI-DI04049098001878

Product

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Reason for recall

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.