Adapt Pump — Class II medical device recall Z-1472-2026
Ongoing recall by Agiliti Health - Ellis, reported 2026-03-04, distributed nationwide. Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support su
| Recall number | Z-1472-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Agiliti Health - Ellis, Ellis, KS, United States |
| Recall initiated | 2026-01-21 |
| Classified | 2026-02-26 |
| Reported by FDA | 2026-03-04 |
| Distributed | Nationwide (US) |
| Quantity | 4,286 |
| Model numbers | 61600200-ADAPT |
Product
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Reason for recall
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.