Adapt Pump — Class II medical device recall Z-1472-2026

Ongoing recall by Agiliti Health - Ellis, reported 2026-03-04, distributed nationwide. Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support su

Recall numberZ-1472-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAgiliti Health - Ellis, Ellis, KS, United States
Recall initiated2026-01-21
Classified2026-02-26
Reported by FDA2026-03-04
DistributedNationwide (US)
Quantity4,286
Model numbers61600200-ADAPT

Product

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Reason for recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.