Data Foundry

Medical device recalls 2020

QIAGEN Rotor Gene Q Software version 2 — Class II medical device recall Z-1476-2020

Terminated recall by Qiagen Sciences LLC, reported 2020-03-18, distributed nationwide. When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the ca

Recall numberZ-1476-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmQiagen Sciences LLC, Germantown, MD, United States
Recall initiated2020-01-28
Classified2020-03-12
Reported by FDA2020-03-18
Terminated2024-09-26
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VE, VN, ZA
Quantity11,418 instruments
Reason classessoftware

Product

QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays.

Reason for recall

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment