cobas pro integrated solutions with cobas c 503 analytical units:… — Class II medical device recall Z-1476-2026
Ongoing recall by Roche Diagnostics Operations, Inc., reported 2026-03-11, distributed nationwide. Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assa
| Recall number | Z-1476-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Operations, Inc., Indianapolis, IN, United States |
| Recall initiated | 2026-01-20 |
| Classified | 2026-02-27 |
| Reported by FDA | 2026-03-11 |
| Distributed | Nationwide (US) |
| Quantity | 1,261 total |
| Reason classes | potency, software |
| UPC / GTIN / UDI-DI | 07613336158852, 07613336179499 |
Product
cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL
Reason for recall
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.