Philips ProxiDiagnost N90 1 — Class II medical device recall Z-1477-2023

Ongoing recall by Philips North America Llc, reported 2023-05-10, distributed nationwide. Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located u

Recall numberZ-1477-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2022-10-21
Classified2023-04-28
Reported by FDA2023-05-10
DistributedNationwide (US)
Countries outside the USAU, CL, CN, DE, ES, IN, IT, MY, NL, PA, TH
Reason classespackaging
UPC / GTIN / UDI-DI00884838085619, 00884838103221
Serial numbers10000000, 10000001, 10000004, 10000005, 10000006, 10000008, 10000009, 10000010, 10000011, 10000012, 10000013, 10000014, 10000015, 10000016, 10000017, 10000018, 10000019, 10000020, 10000021, 10000022, 10000023, 10000024, 10000025, 10000026, 10000027 and 208 more
Model numbers706100, 706110

Product

Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and fluoroscopy (R/F) system

Reason for recall

Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1477-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.