stryker Infinity Resection Adjustment Block — Class II medical device recall Z-1477-2024
Ongoing recall by Wright Medical Technology, Inc., reported 2024-04-10, distributed in NE, TX. Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The
| Recall number | Z-1477-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Wright Medical Technology, Inc., Memphis, TN, United States |
| Recall initiated | 2024-03-06 |
| Classified | 2024-04-04 |
| Reported by FDA | 2024-04-10 |
| Distributed | NE, TX |
| Countries outside the US | AU, BE, BR, CA, CH, CN, CO, DE, ES, FR, GB, IT, SE |
| Quantity | 47 units |
| UPC / GTIN / UDI-DI | 00889797004008 |
| Lot numbers | 2656950, 2762126, 2796094 |
| Model numbers | 33600030 |
Product
stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty
Reason for recall
Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.