RayStation Software Version Numbers version numbers 8B — Class II medical device recall Z-1482-2024

Ongoing recall by RAYSEARCH LABORATORIES AB, reported 2024-04-17, distributed nationwide. A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrec

Recall numberZ-1482-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden
Recall initiated2024-02-23
Classified2024-04-05
Reported by FDA2024-04-17
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, BH, CA, CH, CN, CO, CY, CZ, DE, DK, EE, EG, ES, FR, GB, HK, HU, ID, IL, IN, IT, JP, KR, MA, MY, NL, NO, NZ, PL, QA, SE, SG, SI, TH, TR, TW, ZA
Quantity154 units
UPC / GTIN / UDI-DI07350002010129, 07350002010174, 07350002010204, 07350002010235, 07350002010266, 07350002010297, 07350002010303, 07350002010310, 07350002010365, 07350002010389, 07350002010426, 07350002010433, 07350002010440, 07350002010471, 07350002010488, 07350002010495, 07350002010501, 07350002010549, 07350002010556, 07350002010570, 07350002010587, 07350002010600, 07350002010617, 07350002010631, 07350002010655 and 6 more

Product

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

Reason for recall

A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.