Data Foundry

Medical device recalls 2024

Bed Assist Bars — Class II medical device recall Z-1488-2024

Ongoing recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2024-04-17, distributed nationwide. Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory c

Recall numberZ-1488-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2024-03-05
Classified2024-04-05
Reported by FDA2024-04-17
DistributedNationwide (US)
Countries outside the USCA
Quantity79 devices
Lot numbers15123110004, 15123110005, 15123120001, 15124010001, 15124020002
Model numbersMDS6800SWIV, MDS6800SWIVRH

Product

Bed Assist Bars, labeled as follows: a) Medline SwivAssist Bar, Model Number MDS6800SWIV (case of 3), b) Medline SwivAssist Bar, Model Number MDS6800SWIVRH (individual)

Reason for recall

Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory compliance requirement (ASTM F3186-17, Standard Specification for Adult Portable Bed Rails and Related Products and 16 CFR Part 1270, Safety Standard for Adult Portable Bed Rails) and failed to meet all mandatory requirements outlined in these standards.