MultiDiagnost Eleva — Class II medical device recall Z-1489-2020

Terminated recall by Philips North America, LLC, reported 2020-03-25, distributed nationwide. A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a sh

Recall numberZ-1489-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, LLC, Andover, MA, United States
Recall initiated2019-12-30
Classified2020-03-13
Reported by FDA2020-03-25
Terminated2022-11-17
DistributedNationwide (US)
Quantity534
Reason classesdevice malfunction
Serial numbers01H034592, 01H042957/78, 01H044355, 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 117, 118, 119, 120, 121 and 155 more

Product

MultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038

Reason for recall

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.