MultiDiagnost Eleva — Class II medical device recall Z-1489-2020
Terminated recall by Philips North America, LLC, reported 2020-03-25, distributed nationwide. A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a sh
| Recall number | Z-1489-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips North America, LLC, Andover, MA, United States |
| Recall initiated | 2019-12-30 |
| Classified | 2020-03-13 |
| Reported by FDA | 2020-03-25 |
| Terminated | 2022-11-17 |
| Distributed | Nationwide (US) |
| Quantity | 534 |
| Reason classes | device malfunction |
| Serial numbers | 01H034592, 01H042957/78, 01H044355, 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 117, 118, 119, 120, 121 and 155 more |
Product
MultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038
Reason for recall
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.