Stryker Instruments 0400-820-000S7 T4 XXL Toga w/T3 Cut — Class II medical device recall Z-1498-2013
Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-06-12, distributed nationwide. Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach
| Recall number | Z-1498-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation, Portage, MI, United States |
| Recall initiated | 2013-05-23 |
| Classified | 2013-06-06 |
| Reported by FDA | 2013-06-12 |
| Terminated | 2014-04-14 |
| Distributed | Nationwide (US) |
| Reason classes | sterility |
Product
Stryker Instruments 0400-820-000S7 T4 XXL Toga w/T3 Cut, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
Reason for recall
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1498-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.