Data Foundry

Medical device recalls 2013

Stryker Instruments 0400-830-000 T4 Toga — Class II medical device recall Z-1500-2013

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-06-12, distributed nationwide. Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach

Recall numberZ-1500-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2013-05-23
Classified2013-06-06
Reported by FDA2013-06-12
Terminated2014-04-14
DistributedNationwide (US)
Reason classessterility

Product

Stryker Instruments 0400-830-000 T4 Toga, Zipper, Regular, Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Reason for recall

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1500-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.