Data Foundry

Medical device recalls 2015

Sentec — Class II medical device recall Z-1501-2015

Terminated recall by SenTec AG, reported 2015-04-29. SenTec is voluntarily replacing V-Sign Sensors 2 due to pre-mature deterioration of a part of the PCO2 measuri

Recall numberZ-1501-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSenTec AG, 4106 Therwil, Switzerland
Recall initiated2015-03-25
Classified2015-04-22
Reported by FDA2015-04-29
Terminated2016-02-05

Product

Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil, Switzerland. The SenTec Digital Monitoring System (SDMS) is used for continuous and non-invasive monitoring of transcutaneous carbon dioxide tension (tcPCO2), oxygen saturation (SpO2), and pulse rate (PR).

Reason for recall

SenTec is voluntarily replacing V-Sign Sensors 2 due to pre-mature deterioration of a part of the PCO2 measuring unit and thus disabling PCO2-monitoring. This means that the affected sensors reach their end-of-life earlier than normally expected requiring the sensor`s replacement. There is no increased risk for the patient or operator, because the aforementioned deterioration results in the sens

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1501-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.