Data Foundry

Medical device recalls 2017

Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe… — Class II medical device recall Z-1505-2017

Terminated recall by Ohmeda Medical, reported 2017-03-29, distributed nationwide. Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referr

Recall numberZ-1505-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOhmeda Medical, Laurel, MD, United States
Recall initiated2016-11-18
Classified2017-03-20
Reported by FDA2017-03-29
Terminated2018-07-13
DistributedNationwide (US)
Countries outside the USBR, BS, CA, CO, CR, EC, ES, GB, HN, IE, JP, LB, MX, NL, PA, SV, TT, TW, VE
Reason classesdevice malfunction

Product

Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment for the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Reason for recall

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.