Data Foundry

Medical device recalls 2024

EZ2 Connect MDx-IVD Designed to perform automated isolation and… — Class II medical device recall Z-1508-2024

Ongoing recall by Qiagen Sciences LLC, reported 2024-04-17, distributed nationwide. Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER U

Recall numberZ-1508-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmQiagen Sciences LLC, Germantown, MD, United States
Recall initiated2024-03-04
Classified2024-04-10
Reported by FDA2024-04-17
DistributedNationwide (US)
UPC / GTIN / UDI-DI04053228040576
Model numbers9003230

Product

EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230

Reason for recall

Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1508-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.