IWS Infant Warmer System-Infant radiant warmers provide infrared heat — Class II medical device recall Z-1511-2017

Terminated recall by Ohmeda Medical, reported 2017-03-29, distributed nationwide. Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referr

Recall numberZ-1511-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOhmeda Medical, Laurel, MD, United States
Recall initiated2016-11-18
Classified2017-03-20
Reported by FDA2017-03-29
Terminated2018-07-13
DistributedNationwide (US)
Countries outside the USBR, BS, CA, CO, CR, EC, ES, GB, HN, IE, JP, LB, MX, NL, PA, SV, TT, TW, VE
Reason classesdevice malfunction

Product

IWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled manner, to neonates who are unable to thermoregulate based on their own physiology or necessitate external heat to smoothen the transition from the uterus to the external environment. They fulfill the same purpose of neonatal incubators but they may be preferred over incubators when total access to the patient is needed or desirable (e.g., surgical procedures, extracorporeal membrane oxygenation, frequent resuscitation, etc.). Most infant radiant warmers can be used in two operating modes: 1. Manual: The clinician sets the appropriate heater output for maintaining the desired patient temperature. The heater output is initially selected based on the clinician s training and experience and then is adjusted based on the patients needs and clinical status. 2. Servo: The clinician sets the desired patient temperature. A skin temperature probe senses the patient temperature and feeds this information to the controller of the infant radiant warmer. The controller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Infant radiant warmers have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Infant radiant warmers may incorporate other features, such as phototherapy, observation light, tilting of the bed, and data output to remote monitors or nurse call systems

Reason for recall

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.