Service manuals for the CT — Class II medical device recall Z-1511-2022
Ongoing recall by GE Healthcare, LLC, reported 2022-08-17, distributed nationwide. As a result of unclear guidance in the service manual, the table can drop during a specific service activity i
| Recall number | Z-1511-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2022-06-02 |
| Classified | 2022-08-10 |
| Reported by FDA | 2022-08-17 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, UA, UY, VE, VN, ZA |
| Quantity | 24,843 tables |
| UPC / GTIN / UDI-DI | 00195278098603, 00195278385215, 00195278403087, 00195278420893, 00195278534811, 00195278563408, 00195278568373, 00840682102490, 00840682102551, 00840682102568, 00840682102995, 00840682103053, 00840682105286, 00840682108218, 00840682109796, 00840682116510, 00840682118699, 00840682120975, 00840682125574, 00840682125703, 00840682135313, 00840682137553, 00840682138963, 00840682143523, 00840682143981 and 7 more |
Product
Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.
Reason for recall
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1511-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.