Data Foundry

Medical device recalls 2022

Service manuals for the CT — Class II medical device recall Z-1511-2022

Ongoing recall by GE Healthcare, LLC, reported 2022-08-17, distributed nationwide. As a result of unclear guidance in the service manual, the table can drop during a specific service activity i

Recall numberZ-1511-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2022-06-02
Classified2022-08-10
Reported by FDA2022-08-17
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, UA, UY, VE, VN, ZA
Quantity24,843 tables
UPC / GTIN / UDI-DI00195278098603, 00195278385215, 00195278403087, 00195278420893, 00195278534811, 00195278563408, 00195278568373, 00840682102490, 00840682102551, 00840682102568, 00840682102995, 00840682103053, 00840682105286, 00840682108218, 00840682109796, 00840682116510, 00840682118699, 00840682120975, 00840682125574, 00840682125703, 00840682135313, 00840682137553, 00840682138963, 00840682143523, 00840682143981 and 7 more

Product

Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.

Reason for recall

As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.