GemStar Docking Station — Class I medical device recall Z-1512-2014
Terminated recall by Hospira Inc., reported 2014-05-07, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. There are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in co
| Recall number | Z-1512-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hospira Inc., Lake Forest, IL, United States |
| Recall initiated | 2014-02-18 |
| Classified | 2014-05-01 |
| Reported by FDA | 2014-05-07 |
| Terminated | 2017-01-23 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Countries outside the US | AE, AT, AU, BE, BH, CA, CH, CL, DE, EG, ES, FR, GB, GR, HK, IT, JO, KW, LB, NL, NZ, OM, PT, SA, SI, TR, ZA |
| Quantity | 26,765 docking |
Product
GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and patient bolus cord. The docking station is also used for hosting the GemStar pump for general infusions. List Numbers: 13075-XX-XX.
Reason for recall
There are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List 13000-XX, 13100-XX, or 13150-XX) the potential exists for the GemStar Phase 3 pump to fail to power up while connected to the Docking Station, and 2) when either a GemStar Phase 3 (List 13000-XX, 13100-XX, or 13150-XX)or GemStar Phase 4 pump (List 13086-XX, 13087-XX, or 13088-XX) is used in conjunction with both a Docking Station and an External Battery Pack accessory (List 13073-XX) there exists a possibility that the GemStar pump will display error code 11/003 and give an audible alarm indicating excessive input voltage from the external sources. If the GemStar pump detects what is perceived to be more than 3.6 Volts measured on the external voltage input, the pump will stop the infusion and alarm both with an audible sound as well as a visual alarm.