CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical… — Class II medical device recall Z-1513-2015
Terminated recall by CSA Medical, reported 2015-04-29, distributed nationwide. TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,s
| Recall number | Z-1513-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | CSA Medical, Lexington, MA, United States |
| Recall initiated | 2015-03-25 |
| Classified | 2015-04-23 |
| Reported by FDA | 2015-04-29 |
| Terminated | 2015-08-21 |
| Distributed | Nationwide (US) |
| Quantity | 82 units |
| Serial numbers | 01-00106 |
| Model numbers | CC3-01 |
Product
CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01
Reason for recall
TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax