The Acumed Tension Band Pin is a fixation pin which is — Class II medical device recall Z-1516-2014

Terminated recall by Acumed LLC, reported 2014-05-07, distributed nationwide. Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured w

Recall numberZ-1516-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAcumed LLC, Hillsboro, OR, United States
Recall initiated2014-03-26
Classified2014-04-25
Reported by FDA2014-05-07
Terminated2015-02-04
DistributedNationwide (US)
Countries outside the USAU, CN, ES, FR, GB, IT, JP, KR, SE, ZA
Lot numbers210290, 211349, 214574, 215053, 215405, 216769, 217002, 217004, 217005, 217941, 218273, 218600, 218700, 219145, 219230, 219402, 220978, 221465, 221503, 222731, 222732, 223007, 223008, 223009, 223477, 223756, 224066, 224312, 225554, 225555, 227193, 228951, 229394, 229889, 230287, 230952, 231573, 231971, 232575, 233742 and 37 more
Model numbers30-0097, 30-0098, 30-0099, 50.0, 70.0, 90.0, TB-1550K-S, TB-1570K-S, TB-1590K-S

Product

The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction with commonly used orthopedic wire. 50.0 mm Tension Band Pin; Part Number 30-0097 70.0 mm Tension Band Pin; Part Number 30-0098 90.0 mm Tension Band Pin; Part Number 30-0099 50.0 mm Tension Band Kit; Part Number TB-1550K-S 70.0 mm Tension Band Kit; Part Number TB-1570K-S 90.0 mm Tension Band Kit; Part Number TB-1590K-S Product Usage: Acumed Tension Band Pin is intended to be used with orthopedic wire to address malleolar, patella, and olecranon fractures in tension band wiring procedures. This device is not intended for usage in the spine.

Reason for recall

Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications.