Data Foundry

Medical device recalls 2014

VITROS Chemistry Products FS Diluent Pack 3 — Class II medical device recall Z-1520-2014

Terminated recall by Ortho-Clinical Diagnostics, reported 2014-05-07, distributed nationwide. Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) resu

Recall numberZ-1520-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2014-03-26
Classified2014-04-28
Reported by FDA2014-05-07
Terminated2017-03-27
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Lot numbers01-3266
Model numbers1754, 680
Expiration dates2014-09-23

Product

VITROS Chemistry Products FS Diluent Pack 3, REF 680 1754, 3 x 45 mL, (D1)/15 mL (D2), IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY, 14626

Reason for recall

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1520-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.