Data Foundry

Medical device recalls 2020

Custom procedure trays containing medical components needed to… — Class II medical device recall Z-1520-2020

Terminated recall by Avid Medical, Inc., reported 2020-03-25, distributed in AZ, CA, IL, IN, MD, MS, NJ, NV…. The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locati

Recall numberZ-1520-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAvid Medical, Inc., Toano, VA, United States
Recall initiated2020-02-18
Classified2020-03-19
Reported by FDA2020-03-25
Terminated2020-11-23
DistributedAZ, CA, IL, IN, MD, MS, NJ, NV, NY, TN, TX, VA, WI
Quantity1,001 units
Reason classessterility
Lot numbers1320621, 1329060, 1337288, 1337499, 1340260, 1342539, 1343857, 1344733, 1356016, 1356568, 1357442, 1357815, 1359042, 1362293, 1362314, 1363219, 1363243, 1363557, 1363636, 1363778, 1365656, 1368746, 1370233, 1370391, 1371692, 1372029, 1373730, 1375786, 1376609, 1377795, 1378737, 1379931, 1380256, 1381718, 1381720, 1381721, 1384785
Model numbersADBR013-17, ADCO013-17, ADCO015-17, ADCO016-17, ADVG010-24, ADVH032-25, ADVH049-17, ADVM015-26, ADVM018-25, ADVR034-09, ADVU020-17, NWHR010-01, SINA212
Expiration dates2020-06-30, 2020-12-15

Product

Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.

Reason for recall

The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1520-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.