Software Version 2 — Class II medical device recall Z-1521-2015
Terminated recall by Ortho-Clinical Diagnostics, reported 2015-05-06, distributed in PR. Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended ca
| Recall number | Z-1521-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2015-03-10 |
| Classified | 2015-04-24 |
| Reported by FDA | 2015-05-06 |
| Terminated | 2018-06-01 |
| Distributed | PR |
| Countries outside the US | AU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE |
| Reason classes | potency, software |
| UPC / GTIN / UDI-DI | 10758750001132, 10758750001644 |
| Serial numbers | 34000119, 34002323 |
| Model numbers | 10758750001132, 10758750001644, 6801375, 6801890 |
Product
Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.
Reason for recall
Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).