Data Foundry

Medical device recalls 2015

Software Version 2 — Class II medical device recall Z-1521-2015

Terminated recall by Ortho-Clinical Diagnostics, reported 2015-05-06, distributed in PR. Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended ca

Recall numberZ-1521-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2015-03-10
Classified2015-04-24
Reported by FDA2015-05-06
Terminated2018-06-01
DistributedPR
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Reason classespotency, software
UPC / GTIN / UDI-DI10758750001132, 10758750001644
Serial numbers34000119, 34002323
Model numbers10758750001132, 10758750001644, 6801375, 6801890

Product

Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.

Reason for recall

Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).