Data Foundry

Medical device recalls 2016

Cervical Spine Locking Plate — Class II medical device recall Z-1524-2016

Terminated recall by Synthes (USA) Products LLC, reported 2016-04-27, distributed nationwide. DePuy Synthes is initiating a Voluntary Medical Device Recall of lot # 9666776 and 9671879 of Cancellous and C

Recall numberZ-1524-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSynthes (USA) Products LLC, West Chester, PA, United States
Recall initiated2016-01-06
Classified2016-04-19
Reported by FDA2016-04-27
Terminated2017-01-31
DistributedNationwide (US)
Countries outside the USCA
Lot numbers4MM, 9666776, 9671879
Model numbers450.016, 450.136

Product

Cervical Spine Locking Plate (CSLP) System, 4mm Cancellous Self Tapping Screw, Length 16mm Lot # 9666776 and 4mm Cortex Standard Screw, Length 16mm Lot # 9671879

Reason for recall

DePuy Synthes is initiating a Voluntary Medical Device Recall of lot # 9666776 and 9671879 of Cancellous and Cortex 4.0 mm Screws, which are a part of the Cervical Spine Locking Plate (CSLP) System due to discovery of a single lot of 450.136 screws (4mm Cancellous Self Tapping Screw, Length 16mm, lot 9666776) being swapped with a single lot of 450.016 (4mm Cortex Standard Screw, length 16 mm, lot 9671879).