Medtronic CareLink Encore 29901 Programmer — Class II medical device recall Z-1525-2020

Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2020-03-25, distributed nationwide. Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access

Recall numberZ-1525-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States
Recall initiated2020-01-30
Classified2020-03-19
Reported by FDA2020-03-25
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, AW, BB, BD, BE, BG, BM, BO, BR, BS, CA, CH, CL, CN, CO, CR, CW, CY, CZ, DE, DK, DO, DZ, EC, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, KY, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity5,540 devices
UPC / GTIN / UDI-DI00613994909558, 00643169121973, 00643169166981, 00643169166998, 00643169167025, 00643169167070, 00643169167087, 00643169167094, 00643169225732, 00643169369030, 00643169369047, 00643169763548, 00643169763555, 00643169763562, 00643169763593, 00643169763609, 00643169763623, 00643169955813, 00763000050986, 00763000156640, 00763000156657, 00763000156664, 00763000156671, 00763000156701
Lot numbers00643169, 763593
Model numbers29901, 29901A

Product

Medtronic CareLink Encore 29901 Programmer. Used to interrogate and program Medtronic and Vitatron implantable devices, such as pacemaker/ICD/CRT.

Reason for recall

Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.

Machine access