Data Foundry

Medical device recalls 2024

Yueh Centesis Disposable Catheter Needle — Class II medical device recall Z-1527-2024

Ongoing recall by Cook Incorporated, reported 2024-04-24, distributed nationwide. Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility

Recall numberZ-1527-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCook Incorporated, Bloomington, IN, United States
Recall initiated2024-03-01
Classified2024-04-12
Reported by FDA2024-04-24
DistributedNationwide (US)
Quantity296 units
Reason classespotency, sterility, packaging
UPC / GTIN / UDI-DI00827002094901
Lot numbers15787864, 15804398
Model numbersDTVN-5.0-19-10.0

Product

Yueh Centesis Disposable Catheter Needle - Intended for use in percutaneous fluid aspirations, such as paracentesis or thoracentesis. REFERENCE PART NUMBER (RPN): DTVN-5.0-19-10.0- YUEH ORDER NUMBER (GPN): G09490

Reason for recall

Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1527-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.