Data Foundry

Medical device recalls 2026

GEM Premier 5000 PAK — Class II medical device recall Z-1530-2026

Ongoing recall by Instrumentation Laboratory, reported 2026-03-18, distributed nationwide. Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an

Recall numberZ-1530-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmInstrumentation Laboratory, Bedford, MA, United States
Recall initiated2026-01-20
Classified2026-03-10
Reported by FDA2026-03-18
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KR, LB, LT, LU, LV, MX, MY, NL, NZ, OM, PA, PE, PH, PL, PT, QA, RO, RS, SA, SE, SG, SI, TH, TN, TR, UY, VN, ZA
Quantity1,126 units
UPC / GTIN / UDI-DI08426950807636
Lot numbers251114J, 251117AC, 251117T, 251118B, 251119B, 251121P, 251124O, 251124Q, 251126E, 251201U, 251201V, 251201W, 251202G, 251203H, 251203L, 251204N, 251204O, 251205L, 251208I, 251208J, 251215Y, 251215Z, 251216V, 251217L, 251217X, 251218AA, 251218Y, 251219G, 251219M, 251219N, 251219S, 251222E, 251223D, 251223I, 251223N, 251223O, 251229J, 251229L, 251229U, 251231E
Model numbers00055407510

Product

GEM Premier 5000 PAK; Part No. 00055407510.

Reason for recall

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.