Data Foundry

Medical device recalls 2020

Karl Storz RHINO-LARYNGO-BRONCHO FIBERSCOPE — Class II medical device recall Z-1532-2020

Ongoing recall by Karl Storz Endoscopy, reported 2020-04-01, distributed in AL, AZ, CA, CO, CT, DE, FL, GA…. During a complaint investigation it was determined that an error was observed within the instruction manual in

Recall numberZ-1532-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmKarl Storz Endoscopy, El Segundo, CA, United States
Recall initiated2019-08-27
Classified2020-03-20
Reported by FDA2020-04-01
DistributedAL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV
Quantity21 scopes
Reason classessterility
Model numbers11005BC1

Product

Karl Storz RHINO-LARYNGO-BRONCHO FIBERSCOPE, Model # 11005BC1

Reason for recall

During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.