BARD PERIPHERAL VASCULAR VACCESS¿ PTA Balloon Dilatation Catheters — Class II medical device recall Z-1533-2013
Terminated recall by Bard Peripheral Vascular Inc, reported 2013-06-19, distributed nationwide. Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have
| Recall number | Z-1533-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Bard Peripheral Vascular Inc, Tempe, AZ, United States |
| Recall initiated | 2013-05-28 |
| Classified | 2013-06-12 |
| Reported by FDA | 2013-06-19 |
| Terminated | 2013-10-23 |
| Distributed | Nationwide (US) |
| Quantity | 415 |
| Reason classes | labeling |
Product
BARD PERIPHERAL VASCULAR VACCESS¿ PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess¿ PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess¿ PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.
Reason for recall
Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.