9 — Class II medical device recall Z-1536-2017
Terminated recall by Medical Components, Inc dba MedComp, reported 2017-03-29, distributed in PR. The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit conta
| Recall number | Z-1536-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medical Components, Inc dba MedComp, Harleysville, PA, United States |
| Recall initiated | 2017-02-23 |
| Classified | 2017-03-22 |
| Reported by FDA | 2017-03-29 |
| Terminated | 2018-02-02 |
| Distributed | PR |
| Countries outside the US | PA |
| Quantity | 71 units |
| Reason classes | labeling |
| UPC / GTIN / UDI-DI | 00884908031232 |
| Lot numbers | MHWQ060 |
Product
9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media
Reason for recall
The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer during the insertion procedure.