AquaBplus and AquaBplus B2 Water Purification System — Class II medical device recall Z-1536-2021

Terminated recall by Fresenius Medical Care Holdings, Inc., reported 2021-05-19, distributed nationwide. AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed

Recall numberZ-1536-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Holdings, Inc., Waltham, MA, United States
Recall initiated2021-03-18
Classified2021-05-07
Reported by FDA2021-05-19
Terminated2024-04-10
DistributedNationwide (US)
Quantity27 units
Model numbersG02040109-US

Product

AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. Fresenius P/N 24-2500-0 All 3 units make up the AquaBplus 2500Fresenius Part Number: G02040109-US AquaBplus 3000 - Typ USA G02040115-US AquaBplus 2500 - Typ USA G02040120FMC-US AquaBplus HF- Typ USA

Reason for recall

AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation