Data Foundry

Medical device recalls 2013

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma… — Class II medical device recall Z-1547-2013

Terminated recall by Biomet, Inc., reported 2013-06-26, distributed nationwide. Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Num

Recall numberZ-1547-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBiomet, Inc., Warsaw, IN, United States
Recall initiated2013-04-29
Classified2013-06-14
Reported by FDA2013-06-26
Terminated2015-05-06
DistributedNationwide (US)
Countries outside the USCR, JP, NL
Quantity75
Lot numbers757910
Model numbers14-442021

Product

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

Reason for recall

Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su