Panda¿ iRes Infant Warmers integrated with Nellcor SpO2 — Class II medical device recall Z-1548-2014
Terminated recall by GE Healthcare, reported 2014-05-14. Medical device software may be defective and may cause user-set alarm limits to result in false positive or fa
| Recall number | Z-1548-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, Laurel, MD, United States |
| Recall initiated | 2014-04-02 |
| Classified | 2014-05-06 |
| Reported by FDA | 2014-05-14 |
| Terminated | 2016-05-11 |
| Quantity | 723 units |
| Reason classes | software |
Product
Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
Reason for recall
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.