Data Foundry

Medical device recalls 2026

Owner's Booklets and Instructions for Use that are used with the… — Class I medical device recall Z-1548-2026

Ongoing recall by Trividia Health, Inc., reported 2026-03-18, distributed nationwide. The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act

Recall numberZ-1548-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTrividia Health, Inc., Fort Lauderdale, FL, United States
Recall initiated2026-02-06
Classified2026-03-12
Reported by FDA2026-03-18
DistributedNationwide (US)
Countries outside the USAU, GB, JM, MX, TT
Quantity1,407,465
Reason classeslabeling
UPC / GTIN / UDI-DI00311917178691, 00708820721098

Product

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK

Reason for recall

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1548-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.