HealthCast "Vital Sync" Remote Patient Monitoring System which… — Class II medical device recall Z-1557-2026

Ongoing recall by Covidien LLC, reported 2026-03-18, distributed nationwide. Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alar

Recall numberZ-1557-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCovidien LLC, Boulder, CO, United States
Recall initiated2026-01-30
Classified2026-03-11
Reported by FDA2026-03-18
DistributedNationwide (US)
Countries outside the USES
Quantity78 systems
UPC / GTIN / UDI-DI10884521844933, 10884521844940, 10884521847569, 10884521849358, 10884521856127, 10884521856134

Product

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).

Reason for recall

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.