Data Foundry

Medical device recalls 2016

SOMATOM Force — Class II medical device recall Z-1558-2016

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2016-05-04, distributed nationwide. The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misd

Recall numberZ-1558-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2016-04-01
Classified2016-04-22
Reported by FDA2016-05-04
Terminated2017-04-17
DistributedNationwide (US)
Quantity37 ctsystems
Reason classessoftware
Serial numbers75437, 75439, 75450, 75454, 75458, 75460, 75467, 75475, 75476, 75478, 75481, 75482, 75487, 75493, 75500, 75501, 75510, 75512, 75513, 75514, 75515, 75524, 75526, 75527, 75528 and 12 more

Product

SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.

Reason for recall

The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1558-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.