Echelon — Class II medical device recall Z-1559-2013
Terminated recall by Hitachi Medical Systems America Inc, reported 2013-06-26, distributed nationwide. Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing ima
| Recall number | Z-1559-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hitachi Medical Systems America Inc, Twinsburg, OH, United States |
| Recall initiated | 2013-01-31 |
| Classified | 2013-06-19 |
| Reported by FDA | 2013-06-26 |
| Terminated | 2013-09-20 |
| Distributed | Nationwide (US) |
| Countries outside the US | MX |
| Quantity | 45 systems |
| Reason classes | software |
Product
Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Reason for recall
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.