Storage and Retrieval Module — Class II medical device recall Z-1560-2023
Ongoing recall by Inpeco S.A., reported 2023-05-10, distributed nationwide. Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodiu
| Recall number | Z-1560-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Inpeco S.A., Lugano, Switzerland |
| Recall initiated | 2023-03-10 |
| Classified | 2023-05-04 |
| Reported by FDA | 2023-05-10 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, AU, BE, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EG, ES, FR, GB, GR, HU, IE, IL, IN, IT, JP, KE, KR, MX, MY, NL, NO, OM, PK, PT, QA, SA, SE, SG, SI, SK, TH, TR, TW, VN, ZA |
| Reason classes | device malfunction, software |
| UPC / GTIN / UDI-DI | 07640172340004, 07640172341001, 07640172342008 |
| Model numbers | ACP-207-00, ACP-207-01, ACP-230-00, ACP-230-01, APT-207-00, APT-207-01, APT-230-00, APT-230-01, FLX-207-00, FLX-207-01, FLX-207-02, FLX-207-10, FLX-207-11, FLX-230-00, FLX-230-01, FLX-230-02, FLX-230-10 |
Product
Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: ACP-207-00, ACP-207-01, ACP-230-00, ACP-230-01 APT-207-00, APT-207-01, APT-230-00, APT-230-01 FLX-207-00, FLX-207-01, FLX-207-02, FLX-230-00, FLX-230-01, FLX-230-02 FLX-207-10, FLX-207-11, FLX-230-10
Reason for recall
Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1560-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.