Data Foundry

Medical device recalls 2026

Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN — Class II medical device recall Z-1563-2026

Ongoing recall by Medline Industries, LP, reported 2026-03-18, distributed nationwide. Fabric may experience premature delamination when using included laundering instructions. Identifying delamina

Recall numberZ-1563-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2026-01-28
Classified2026-03-12
Reported by FDA2026-03-18
DistributedNationwide (US)
Countries outside the USAE, CA, CL, IN, PA, SA
Quantity104,004
UPC / GTIN / UDI-DI10193489021820, 10193489040111, 10197344014002, 10884389189191, 10884389189221, 10884389193297, 10888277189188, 10888277189218, 20193489021827, 20193489040118, 20197344014009, 40884389189192, 40884389189222, 40884389193298, 40888277189189, 40888277189219

Product

Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL

Reason for recall

Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1563-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.