Data Foundry

Medical device recalls 2014

ABACUS TPN Calculation Software — Class I medical device recall Z-1566-2014

Terminated recall by Baxter Corporation Englewood, reported 2014-05-21, distributed nationwide. 1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creatio

Recall numberZ-1566-2014
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Corporation Englewood, Englewood, CO, United States
Recall initiated2014-04-02
Classified2014-05-12
Reported by FDA2014-05-21
Terminated2016-01-20
DistributedNationwide (US)
Countries outside the USAE, AT, BG, BH, CA, CH, CO, DE, DO, EE, FR, GB, HK, HU, IE, IL, JO, KW, LV, MA, OM, PA, QA, SA, SI, SK, TW, ZA
Quantity2,034
Reason classessterility, software

Product

ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition), 8300-0156 ABACUS 3.0 CE, 8300-0157 ABACUS 3.0 SE, 8300-0158 ABACUS 3.0 ME, 8300-0166 ABACUS 3.1 DE, 8300-0167 ABACUS 3.1 CE, 8300-0168 ABACUS 3.1 SE, 8300-0169 ABACUS 3.1 ME. ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for Total Parenteral Nutrition (TPN) order calculations.

Reason for recall

1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1566-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.