Medtronic SynchroMed¿ II Implantable Drug Infusion Pump — Class II medical device recall Z-1570-2014
Terminated recall by Medtronic Neuromodulation, reported 2014-05-14. This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed
| Recall number | Z-1570-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2014-02-26 |
| Classified | 2014-05-08 |
| Reported by FDA | 2014-05-14 |
| Terminated | 2018-09-28 |
| Model numbers | 8637-20, 8637-40 |
Product
Medtronic SynchroMed¿ II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.
Reason for recall
This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1570-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.