Data Foundry

Medical device recalls 2014

Medtronic SynchroMed¿ II Implantable Drug Infusion Pump — Class II medical device recall Z-1570-2014

Terminated recall by Medtronic Neuromodulation, reported 2014-05-14. This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed

Recall numberZ-1570-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Neuromodulation, Minneapolis, MN, United States
Recall initiated2014-02-26
Classified2014-05-08
Reported by FDA2014-05-14
Terminated2018-09-28
Model numbers8637-20, 8637-40

Product

Medtronic SynchroMed¿ II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.

Reason for recall

This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re