Data Foundry

Medical device recalls 2016

BD Trucount Absolute Counting Tubes: Model number: 340334 IVD — Class II medical device recall Z-1570-2016

Terminated recall by Becton, Dickinson and Company, BD Biosciences, reported 2016-05-04, distributed in CA, CO, CT, DE, FL, GA, LA, MD…. One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead cou

Recall numberZ-1570-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBecton, Dickinson and Company, BD Biosciences, San Jose, CA, United States
Recall initiated2016-03-31
Classified2016-04-26
Reported by FDA2016-05-04
Terminated2017-05-26
DistributedCA, CO, CT, DE, FL, GA, LA, MD, MN, NC, NH, NY, OH, PA, TN, VA, WA
Countries outside the USCN, IN, JP
Lot numbers534964
Model numbers340334, 340334IVD

Product

BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.

Reason for recall

One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.