ACUSON X700 Ultrasound System with software version 1 — Class II medical device recall Z-1572-2016
Terminated recall by Siemens Medical Solutions USA, Inc., reported 2016-05-04. Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasou
| Recall number | Z-1572-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Medical Solutions USA, Inc., Mountain View, CA, United States |
| Recall initiated | 2015-09-05 |
| Classified | 2016-04-27 |
| Reported by FDA | 2016-05-04 |
| Terminated | 2016-04-27 |
| Countries outside the US | AE, BR, DE, HU |
| Quantity | 4 |
| Reason classes | software |
| Serial numbers | 10658844 |
| Model numbers | 10658844 |
Product
ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Reason for recall
Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1572-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.