MICS3 Angled Sagittal Saw Attachment — Class II medical device recall Z-1572-2026
Ongoing recall by Howmedica Osteonics Corp., reported 2026-03-25, distributed nationwide. A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal
| Recall number | Z-1572-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Howmedica Osteonics Corp., Mahwah, NJ, United States |
| Recall initiated | 2026-02-02 |
| Classified | 2026-03-18 |
| Reported by FDA | 2026-03-25 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, DE, ES, FR, GB, NZ |
| Quantity | 1,098 units |
| Reason classes | potency, device malfunction |
| UPC / GTIN / UDI-DI | 07613327523874 |
| Model numbers | 210490 |
Product
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Reason for recall
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1572-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.