NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body… — Class II medical device recall Z-1573-2016
Terminated recall by Philips and Neusoft Medical Systems Co., Ltd., reported 2016-05-04, distributed in CT, FL, LA, MO, NC, NE, OH, PR…. The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426:
| Recall number | Z-1573-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips and Neusoft Medical Systems Co., Ltd., Shenyang, N/A, China |
| Recall initiated | 2016-03-24 |
| Classified | 2016-04-27 |
| Reported by FDA | 2016-05-04 |
| Terminated | 2016-11-23 |
| Distributed | CT, FL, LA, MO, NC, NE, OH, PR, SC, TX |
| Quantity | 37 units |
| Reason classes | software, unapproved product |
| Serial numbers | N16E090002, N16E090003, N16E090007, N16E090008, N16E090010, N16E100017, N16E100023, N16E110023, N16E110026, N16E120016, N16E120023, N16E120052, N16E120053, N16E130012, N16E130038, N16E130045, N16E140007, N16E140008, N16E140012, N16E140017, N16E140020, N16E140029, N16E140030, N16E140031, N16E140056 and 12 more |
Product
NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Reason for recall
The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be