Data Foundry

Medical device recalls 2013

Medtronic Intrathecal Catheter Pump Segment Revision Kit — Class I medical device recall Z-1574-2013

Terminated recall by Medtronic Neuromodulation, reported 2013-07-03, distributed nationwide. The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for o

Recall numberZ-1574-2013
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Neuromodulation, Minneapolis, MN, United States
Recall initiated2013-06-03
Classified2013-06-25
Reported by FDA2013-07-03
Terminated2014-04-07
DistributedNationwide (US)
Countries outside the USAE, AT, AU, AW, BE, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, EC, ES, FI, FR, GB, GR, HK, HU, IE, IL, IN, IS, IT, JO, KR, KW, LB, LK, LT, LU, MA, MT, NL, NO, NZ, PA, PL, PT, QA, RO, SA, SE, SG, SI, SK, TH, TR, TT, UY, VE, ZA
Quantity115,722 total
Reason classesdevice malfunction
Model numbers8596SC, 8731, 8731SC

Product

Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC. Contents: 60-cm pump segment with attached sutureless pump connector, Spinal Segment Strain-relief sleeves, Pump segment strain-relief sleeves, Connector pin. Contents of inner package are STERILE. The Medtronic Model 8596SC Pump Segment Revision Kit is used when a revision to the pump segment of the Model 8731 or Model 8731SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8731 or Model 8731SC catheter. The catheter connects to the pump at the catheter port.

Reason for recall

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1574-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.